Surface contact / tube contact temperature measuring transducer THERMASGARD® ALTM2-I Q LCD

€195.11
Tax included

Calibratable THERMASGARD® ALTM 2 tube contact temperature measuring transducer with eight switchable measuring ranges, external sensor, continuous output, housing made of impact resistant plastic with quick-release screws, optionally with/without display, with cable gland or M12 connector according to DIN EN 61076-2-101.

The surface-contact sensor device is used for temperature detection on lines, pipes (e.g. cold and hot water) or on heating sections for heating system control.

THERMASGARD®
Protection type
Output model
Output
Display
Quantity

Power consumption

< 1,0 VA/ 24 V DC

< 2,2 VA/ 24 V AC

Voltage supply

15...36 V DC

depending on working resistance, residual ripple stabilised ± 0.3 V

Temperature deviation

typically ± 0.2 K at +25 °C

Working resistance

Ra (ohm) = (Ub -14 V)/ 0.02 A for I variant

Sensor

Pt1000, DIN EN 60751, class B

Connection type

2- or 3-wire connection

Ambient temperature

Measuring transducer –30...+70 °C

Process connection

endless metal strap and metal tightener

(included in scope of delivery)

Insulating resistance

100 MΩ,

at +20 °C (500V DC)

Strap dimensions

Ø = 13 - 92 mm (¼ - 3"), L = 300 mm

Electrical connection

0.14 - 1.5 mm² via terminal screws

Cable gland

M 16 x 1.5, including strain relief, exchangeable,

max. inner diameter 10.4 mm

sensor cable

silicone, SiHF, 2 x 0.25 mm², KL = 1.5 m

Output

4...20 mA

Permissible air humidity

< 95 % r. H., non-precipitating air

Protection class

III (according to EN 60 730)

Sensor protection

pipe feeder, stainless steel V4A (1.4571), Ø = 6 mm,

nominal length NL = 50 mm, cable entry stamped

Protection type housing

IP 65 (according to EN 60 529) Housing tested,

TÜV SÜD, Report No.713139052 (Tyr 1)

Measuring ranges

multi-range switching with 8 switchable measuring ranges,

(other ranges optional)

detached sensor variant: Tmax over 100 °C, operating range –50... +150 °C;

with manual zero point correction (± 10 K)

Cable connection

cable gland, plastic (M 16 x 1.5; with strain relief, exchangeable, max. inner diameter 10.4 mm) or

M12 connector according to DIN EN 61076-2-101 (optional on request)

Housing

plastic, UV-stabilised,

material polyamide, 30 % glass-globe reinforced,

with quick-locking screws

(slotted/ Phillips head - combination),

colour traffic white (similar to RAL 9016),

housing cover for display is transparent !

Housing dimensions

72 x 64 x 43.3 mm (Tyr 1 with Display)

Protection type

IP 65 (according to EN 60 529) humidity-tight stamped

IP 68 (optional watertight compound-filled) rolled

Protection type (sensor)

IP 65

2001-2122-2100-001

16 same category products: